If you own the Physiology (Dynamics) module you can use the MeanROI setup to specify user-defined measurement regions (ROIs) before the acquisition of your time lapse experiment and analyze their time­‑dependent changes in intensity online during acquisition. Ratios can also be calculated and displayed online - these are the typical functions used in physiology/ calcium Fura-2 applications.

Before the experiment

A precondition for a physiology experiment is a Time Series experiment (which can include a z-stack acquisition), which is set up in the Time Series tool. Adding a time dimension to your experiment allows you to activate the optional tool Dynamics. This tool contains the button for opening MeanROI Setup. Here you can draw in ROIs and adjust the display layout of the measurement results. Note that when the setup is opened a snap is automatically acquired, on the basis of which you can configure the settings for the subsequent experiment, like for example the ratio parameters. The structure of the MeanROI setup is based on the MeanROI view, making it easier to learn. Note that online measurements on tiles and positions experiments are not possible, but you can perform measurements on tiles or position data collected over time (i.e. multi-scene time series) in MeanROI view post-acquisition.

During the experiment

After being started, physiology experiments are displayed in the online mode of the MeanROI View. This allows you to analyze and follow the experiment during acquisition. The structure and options largely correspond to the offline mode of the MeanROI View. We therefore recommend that you familiarize yourself with the MeanROI View (offline) before performing your Physiology experiment.

After the experiment

After you have performed your Physiology experiment the data are displayed in the offline mode of the MeanROI view and can be analyzed, processed and exported there. For more information, see also Workflow MeanROI View (Offline).

  1. To perform physiology experiments, you need the Physiology (Dynamics) module.
  2. You have created a new experiment, defined at least one channel and adjusted the focus and exposure time, see also Setting Up a New Experiment and Acquiring Multi-Channel Images with Cameras.
  3. You have licensed this functionality and activated it under Tools > Toolkit Manager.
  4. You are on the Acquisition tab.
  1. Activate Time Series in the Acquisition Dimensions section.
  2. The Time Series tool is displayed in the Left Tool Area under Multidimensional Acquisition.
  3. Activate Dynamics in the experiment manager.
  4. The Dynamics tool now is displayed in the in the Left Tool Area under Applications.
  5. Note that the tool is not available if the Tiles or Panorama dimensions are activated. Deactivate these dimensions to make the tool available.
  6. Set up a time series experiment, see Acquiring Time Series Images.
  7. Open the Dynamics tool.
  8. Click MeanROI Setup.
  1. You have completed the general prerequisites for Physiology experiments.