Acquiring Time Series Images
- To set up Time Series experiments, you need to have licensed this functionality and activated it under Tools > Toolkit Manager. For LSM systems, this functionality is part of the system license.
- You have set up a new experiment with at least one defined channel and adjusted focus and exposure time correctly, see also Setting Up a New Experiment and Acquiring Multi-Channel Images with Cameras.
- You are on Acquisition tab.
- In the Acquisition Dimensions section, activate Time Series .
- The Time Series tool is displayed in the Left Tool Area.
- Open the Time Series tool.
- Set the Duration of your time series. You are able to select an interval (days, hours, minutes, seconds, milliseconds) or cycles (1-n), e.g. 10 cycles.
- Set the Interval of your time series.
- Click Start Experiment.
- The time series experiment starts. You have successfully set up a time series experiment.
You can display the individual images via the Time slider on the Dimensions tab.