This method provides an alternative processing of confocal data for even higher resolution. An unprocessed confocal file needs to be selected as an input file.

LSM Deconvolution enables you to adjust the strength of the processing by defining the number of iterations or additionally the quality threshold. Note that overdoing the processing strength can result in artificially small structures, or splitting of solid structures.

Once processing is complete, review the output. Check the image Info view to see the number of iterations actually used, which can be informative especially if you have set a Quality Threshold. Save the processed image file with a new name to preserve the original unprocessed file for potential re-evaluation or different processing attempts. Ensure that the parameters are set according to the specifics of your sample and imaging requirements to avoid artifacts such as artificially small structures or splitting of solid structures.

Parameter

Description

Start with Last Result

Activated: Keeps the iteration result and adds additional iterations instead of starting the processing from the beginning again. Should be done with the same Pixel Upsampling setting.

Pixel Upsampling

Activated: Increases the sampling of output image by the factor of two for the best resolution.

Sample Structure

Sets the default value for typical sample types.

–

Sparse

Sets the default value for a sample with small structures and lots of black background (e.g. beads).

–

Standard

Sets the default value for a standard sample.

–

Dense

Sets the default value for a sample with lots of structures and little background.

Maximum Iterations

Sets the number of maximum iterations. This input is connected to the Quality Threshold parameter. If the Quality Threshold is set to 0, the number of iterations are processed as set here. If a threshold greater 0 is defined, the set number of iterations will not always be used, but processing stops if a certain resolution quality is achieved. The number of actual iterations is documented in the image Info view.

Quality Threshold

Sets the threshold for quality. This input is connected to the Maximum Iterations parameter.