Selecting Autofluorescence Spectra
- You have started the wizard for calibrated unmixing, see Starting the Calibrated Unmixing Wizard.
- You have prepared reference scans of your sample, see Information for Acquisition and Preparation of Reference Scans.
- In the selection field, click
.
- A file browser opens.
- Select the image of your unstained reference scan and click Open. If a spectrum of the autofluorescence is already available from a previous run of this wizard, select the spectra file.
- The file is loaded. The image or the spectrum of the selected file is displayed in the Center Screen Area.
- In case you have loaded an image, go to the Graphics tab below the image.
- The controls for graphical elements are displayed, see Graphics Tab.
- Use the 2D or Split view and the controls of the Graphics tab to draw at least one graphical element which contains a typical sample of the relevant type of autofluorescence.
- You have marked the autofluorescence.
- If a part of the image also contains only background, for example no tissue but empty glass, draw a graphical element into the image to mark the background.
- The relevant information is marked in the image.
- Go to Select Autofluorescence and select the graphical element that marks the autofluorescence.
- Autofluorescence is set.
- Go to Select Background and select the graphical element that marks the background, if available. Otherwise select None from the drop-down list.
- The background is set.
- Click Next.
- The next step opens, see Selecting Dye Spectra.