Displays advanced microscope information that influences the form of the PSF in a channel-dependent way:

Parameter

Description

Illumination

Selects the illumination method with which the data set has been acquired. In the event that a Conventional Microscope has been entered under the microscope parameters, the following options are available here: Epifluorescence, Multiphoton Excitation and Transmitted Light. In the case of confocal microscopes, Epifluorescence is the only option.

Image Formation

Displays whether the imaging was incoherent (Conventional Microscope) or coherent (Laser Scanning Microscope).

Lateral Resolution

Displays the lateral resolution of the calculated PSF.

Axial FWHM

Displays the FWHM (Full Width Half Maximum) as a measure of the axial resolution of the PSF.