The following table lists the parameters which are used by default for widefield, confocal, lightsheet and ApoTome images.

Microscope Type

Widefield

Confocal

Lightsheet

ApoTome

General Deconvolution Parameter Defaults

Normalization

Auto

Auto

Auto

Auto

Background Correction

Off

Off

Off

Off

Flicker Correction

Off

Off

Off

Off

Decay Correction

Off

Off

Off

Off

Hot Pixel Correction

Off

Off

Off

Off

Constrained Iterative Specific Defaults

Strength (automatic, manual; range 0..10)

Auto

Auto

Manual (str 5)

Auto

Likelihood

Poisson

Poisson

Poisson

—

Regularization

Zero Order

First Order

Zero Order

—

Optimization

Analytical

Line Search

Analytical

—

First Estimate

Input

Mean

Input

—

Maximum Number Of Iterations

40

7

40

—

Auto Stop Percentage

0.1

0.1

0.1

—

Fast Iterative Specific Defaults

Method

Poisson/Meinel

Poisson/Richardson Lucy

Poisson/Meinel

—

Regularization

None

None

None

—

Optimization

None

None

None

—

First Estimate

Input

Mean

Input

—

Maximum Number Of Iterations

15

50

15

—

AutoStop Percentage

0.1

0.1

0.1

—

Regularized Inverse Specific Defaults

Regularization

Zero Order

Zero Order

Zero Order

First Order