Configuring a Z-Stack Automatically
If you use an optical-sectioning method (for example LSM and Apotome), an automatic configuration of the z-stack is not available. Configure the z-stack manually with the Z-Stack tool and start the acquisition by clicking Start Experiment.
To change the area of the sample (z-direction in %) covered by the z-stack auto configuration, go to Tools > Options > Acquisition > Z-Stack. Smaller values enlarge the z-stack, higher values make the z-stack smaller.
- The camera's current field of view is always at a position on the sample that shows a signal in the selected channel.
- The current focus position is at the center of the sample.
- The exposure time of the camera is set to receive a good signal.
- On the Acquisition tab, in the Z-Stack tool, click Start Auto Configuration.
- A message is displayed.
- Confirm the system message by clicking OK.
- The automatic configuration starts. The auto-configuration can last for a few seconds up to half a minute, depending on the acquisition settings. You can check the status on the Progress bar in the Status bar.
- The automatic configuration sets the focus position for the first, last, and center slice of the z-stack, the number of slices, and the interval automatically.
- Click Start Experiment.
- You have successfully set up and started a z-stack experiment.