Channel Specific Section
In this section you find all settings that are channel specific. This means that they can be configured differently for each channel.
|
Parameter |
Description |
|---|---|
|
Illumination |
Displays the excitation wavelength for the channel dye [in nm] by using the peak value of the emission spectrum. The color field corresponds to the wavelength (as far as possible). |
|
Detection |
Displays the peak value of the emission wavelength for the channel dye. The color corresponds to the wavelength (as far as possible). |
|
Sampling Lateral |
Depends on the geometric pixel scaling in the X/Y direction and displays the extent of the oversampling according to the Nyquist criterion. The value should be close to 2 or greater in order to achieve good results during DCV. As, in the case of widefield microscopes, this value is generally determined by the objective, the camera adapter used and the camera itself, it can only be influenced by the use of an Optovar. With confocal systems, the zoom can be set to match this criterion. |
|
Sampling Axial |
Depends on the geometric pixel scaling in the Z direction and displays the extent of the oversampling according to the Nyquist criterion. The value should be at least 2 or greater in order to achieve good results during DCV. This value is determined by the increment of the focus drive during acquisition of Z-stacks and can therefore be changed easily. |