Selecting Dye Spectra
- You have selected the autofluorescence spectra and are in the second step of the Calibrated Unmixing wizard, see Selecting Autofluorescence Spectra.
- You have prepared reference scans of your sample, see Information for Acquisition and Preparation of Reference Scans.
- For Unmixing Database Folder, define a name for the dye set to be used in the unmixing library.
- You have defined a name.
- Go to the table in the Center Screen Area, select the first entry, and click
.
- A file browser opens.
- Select one of the (cropped) single-stain reference scan images and click Open.
- The file name is displayed in the table and the image is displayed below the table for the selected entry.
- For ChName, select the channel where the dye is most visible. Use the Split view of the image displayed below the table to identify the correct channel.
- You have selected the channel.
- For Dye, enter a name for the dye for later reference.
- Dye name is set.
- For AF, select the channel which mostly represents pure autofluorescence (i.e. has no specific signal) and has a good signal-to-noise ratio. Usually, the channels suggested by the system can be used. Otherwise, Use the Split view of the image displayed below the table to identify the correct channel.
- The channel for autofluorescence is set.
- Repeat the previous steps with the single-stain reference scans for all dyes, one per row.
- The dye spectra are set.
- Click Next.
- The next step opens, see Checking the Calculated Dye Spectra.