Extended Depth of Focus
With this method, you can combine the sharp regions from the individual sections of a z-stack image to create a single sharp image. This enables you to display a considerably larger depth of field than is possible on a microscope.
|
Parameter |
Description |
|
|---|---|---|
|
Method |
||
|
- |
Wavelets |
Uses a wavelet transformation to detect the sharpest areas in the images. |
|
- |
Contrast |
For this method, the value is the difference between the brightest and the darkest pixel value within the “Kernel”. |
|
- |
Maximum Projection |
Images with the brightest and darkest pixels are generated first. Of these images the image with the higher variance is used as the resulting image. Usually, best suited for fluorescence images. |
|
- |
Variance |
Calculates the variance of the pixel values within the “Kernel”. Usually, best suited for bright field images. |
|
- |
Fast EDF |
Uses a method for fast EDF to detect the sharpest areas in the images. |
|
Z-Stack Alignment |
This parameter should not be used for processing whole slide image scans. It is only required when processing images stemming from a stereo microscope. Not visible if you have selected Maximum Projection as Method. |
|
|
- |
No Alignment |
The z-stack image is not aligned before the calculation. Select this setting if the z-stack image has not been acquired using a stereo microscope. |
|
- |
Normal |
Uses an alignment with high speed for normal image quality. |
|
- |
High |
Uses an alignment with lower speed for high image quality. |
|
- |
Highest |
Uses an alignment with the lowest speed with best image quality. |
|
Process Tiles Separately |
Only visible if you have selected Maximum Projection as Method. |
|
|
Contrast Lengthscale |
Only visible if Contrast or Variance is selected as Method. |
|
|
Smoothing |
Only visible if Contrast, Variance, or Fast EDF is selected as Method. |
|
|
Reconstruction |
Only visible if Contrast or Variance is selected as Method. |
|
|
Decay Flicker Correction |
Only visible if Contrast or Variance is selected as Method. |
|
|
Pixelwise Iteration |
Only visible if Fast EDF is selected. Default values (depending on the selected preset):
|
|
|
Depthwise Kernel Size |
Only visible if Fast EDF is selected. Default values (depending on the selected preset):
|
|
|
Resize Ratio |
Only visible if Fast EDF is selected. Default Value: 0.25
|
|