Image quality
Because tracking is dependent on the ability to recognize objects over time, it is first, and foremost, dependent on the ability to recognize objects. This means that the image quality must be good enough for segmentation algorithms to work, or at the very least for manual identification of objects by the user.
Segmentation itself is the process of identifying which pixels in an image are part of objects, and also of identifying which pixels belong to which object so that touching objects can be separated. The simplest form of image segmentation is known as thresholding. This is where we set a rule that all pixels with an intensity either above or below a specific threshold are classified into the Objects pixel class, and the rest are classed into the Background class. Then we can just say that contiguous groups of pixels are individual objects.
Several aspects of images can make this process more or less difficult. For example, a noisy image will have a lot more punctual variation that could end up with a lot of single-pixel objects or very rough object outlines. Also, some modalities can be particularly better or worse suited for segmentation. Fluorescence images are generally easier to segment compared to phase contrast or DIC images.
This image, for example, is quite challenging for segmentation using thresholds. There are variations of intensity inside what might be considered objects that could cause the head and tail to be separated. The objects themselves are out of focus so their boundaries are imprecise which could lead to difficulties in separating objects in close proximity, and there are big variations in the background intensity which could make a simple threshold difficult with some pre-processing.
The same sample images under different conditions (e.g. darkfield illumination) could make the segmentation comparatively much easier.