Registering using landmarks considers the image spatial calibrations to work out the scaling required (registering using surface transformations ignores calibrations in favor of the user-defined scaling). This is usually not an issue when fusing images captured using the same microscope and the same magnification. However, if the two volumes are imaged using different imaging devices the calibrations could be different and this could lead to the fused volumes having a scaling offset. Also, when doing multimodal imaging where the sample may receive additional preparation steps for the second acquisition, the sample can shrink or grow resulting in a calibration that is technically correct, but effectively wrong. In such cases, we can either correct the calibrations to account for changes in the sample or use the surface transformation to disregard the calibration and adjust the scaling as required.