Seeding

Parameter

Description

Method

Sets the seeding method.

–

Nucleus-based

Detects the bright peaks such as nuclei.

–

Membrane-based

Detects dark peaks inside membrane staining.

Channel

Sets the analysis target channel.

Seeded Detection

Sets the seed threshold percentage. A lower value detects fewer, but bigger seeds. A higher value detects more, but smaller seeds.

Area

Sets the Area, Volume or Diameter seeds range. Only the segments that matches the range are accepted as seeds. The filter can be enabled/disabled.
Note: You can measure the diameter directly from the dataset with the Measurement tool.

–

Diameter

Filters the unwanted seeds based on their diameter.

–

Volume

Filters the unwanted seeds based on their volume.

–

Area

Filters the unwanted seeds based on their area.

Region Growing

Parameter

Description

Method

Sets the region growing method.

–

Watershed

Segments bright objects such as cells by growing from seeds to the border of an object.

–

Membrane-based watershed

Segments dark objects such as membrane staining by growing from the seeds to the bright intensity outlines.

Channel

Sets the channel for the selected algorithm.

Threshold

Sets this parameter so that the foreground of the region growing channel contains the objects of interest.

Max. distance

Sets the maximum distance on which the region growing will act.
Note: You can measure the distance directly from the dataset with the Measurement tool.