ZEISS Microscopy Knowledge Base
Help
ZEISS Microscopy Knowledge Base

Input ROI

This operation allows to select the region of interest (ROI). ROI defines the dataset subarea that will be processed and analyzed by the pipeline.

Input ROI dialog: ROI Current time point; Channels BiG R 1; Scaling 100% (Include Z checked); Crop input data unchecked

Input ROI parameters

Parameter

Description

ROI

Sets the processing and analysis target space.

Current View

The selected Z plane and the viewer area are processed.

Current Plane

The selected Z plane is processed (XY).

Current Time Point

The selected time point is processed (XYZ).

Current Image Set

The complete dataset (XYZ and time) is processed.

Custom

Allows to mix the previous methods.
Note: Use the Custom option during the pipeline setting and testing. Set a sub volume (XY, Planes, Time Points, channels) of your dataset on which perform the trial. This will speed up the setting process.

Channels

Sets the processing and analysis target channels. Selecting a single channel, all the operations in the pipeline will be forced to use it.

Scaling

Sets the scaling of the dataset, which reduces it size. The measurements will not be modified by the scaling factor

Crop input data

Sets the limitation of the data used for the calculation to only the cropped selection defined above. If this option is not selected, the entire image set is used for the calculation.

Additional Parameters

Additional settings for Input ROI expand, when you click Two dark downward chevrons inside a small light-gray rounded square icon in the operation.

Parameter

Description

Bounds

Sets the analysis area edges. The whole XY bounds, the viewing area or a custom space can be applied.

Planes

Sets the analysis planes range. A single plane, a range of planes or the whole stack can be selected.

Time Points

Sets the analysis time points range. A single time pint, a range of time points or the whole movie can be selected.

Impressum
Carl-Zeiss-Strasse 22
73447 Oberkochen
Germany
Legal