Image registration in a nutshell
Image registration is the process of transforming different sets of data to overlay images from different experiments, taken at different times, positions or angles, or with different imaging modalities. The algorithm attempts to discover the matching areas and align them together. Image registration methods can be divided into two large groups: rigid and non-rigid transformations. For the multiplexing experiments, we will use the nuclei staining in each of the experiment cycles to register the subsequent experiments to the first image. This method is applicable both to 2D and 3D images.
The application workflow starts from importing the original image files (for example, CZI) in arivis Pro and registering the subsequent experimental cycles to the first one based on the indicated nuclei staining channel (i.e. DAPI or Hoechst). The images will be imported as a stack of individual tiles with all the channels from all the experiment cycles, each registered to the nuclei staining in the first experiment cycle. After the registration, the mosaic image should be stitched using the Tile Sorter tool.