Tips to Ensure Accuracy
On the light microscope
- To ensure good calibration, the objectives must be correctly calibrated.
- We recommend 20x magnification for the automatic marker detection.
- Due to potential objective shift, accuracy will be higher if you acquire images using the same objective that you used for calibration.
On the electron microscope
- The SE2 detector is well suited for marker detection.
- Switch off the beam shift and scan rotation.
- The "Stage only" setting should also be activated under Center Point or Feature.
- Use the stage backlash to increase the stage’s accuracy in locating positions.
General Points
- For images on a SEM use conductible samples.
- Initialize the stages before using Shuttle & Find.
- Calibrate again each time the holder has been inserted into a microscope.
- If the marker is a little dirty, it can be helpful for detection purposes to de-focus it slightly.
- In the case of stitched tiles images accuracy is reduced.