On the light microscope

  • To ensure good calibration, the objectives must be correctly calibrated.
  • We recommend 20x magnification for the automatic marker detection.
  • Due to potential objective shift, accuracy will be higher if you acquire images using the same objective that you used for calibration.

On the electron microscope

  • The SE2 detector is well suited for marker detection.
  • Switch off the beam shift and scan rotation.
  • The "Stage only" setting should also be activated under Center Point or Feature.
  • Use the stage backlash to increase the stage’s accuracy in locating positions.

General Points

  • For images on a SEM use conductible samples.
  • Initialize the stages before using Shuttle & Find.
  • Calibrate again each time the holder has been inserted into a microscope.
  • If the marker is a little dirty, it can be helpful for detection purposes to de-focus it slightly.
  • In the case of stitched tiles images accuracy is reduced.