Panorama Acquisition Options Tool
This tool can only be added to the Panorama (automatic) workbench. Note that the options should only be activated, if you have problems with the camera/live image during acquisition. It is only available for troubleshooting up and normally not needed.
The default state of the checkboxes is set to "undefined" (checkboxes are marked with a rectangle) which means the software tries to use an determined value for the current camera automatically. Meaning if you are unsure, if your camera supports the functionality, we recommend to leave the state to "undefined".
|
Parameter |
Description |
|---|---|
|
Use Camera Streaming (if possible) |
If activated, the software uses the free running/streaming mode of the camera. This option is intended for slower cameras. |
|
Stop and Go |
If activated, during acquisition you have to move the stage, e.g. about one camera frame, and the stop the movement. Then you have to move the stage for the next frame and stop again. Compared to the "default" mode, the images to be stitched are acquired only when the stage movement stops. This option is intended for cameras that produce motion artefacts (e.g. rolling shutter effect) and for exposure times greater than about 50 ms. |