Size Distribution View
The Size Distribution view displays the particle results for the selected standard. You can filter the results by approval status, particle type or particle size class.
In the Image Gallery, all particles are displayed from 5 µm length and width onwards. As a Supervisor, you can change the minimum size of the displayed particles in the Region Filter tool in the Region Filter workbench. By default, the Region Filter workbench is set to run silent, see Region Filter.
For individually created job templates it is recommended to display only standards with the same smallest particle size in the Standard Selection tool and adapted the Region Filter tool accordingly.
|
1 |
Class Chart Displays the number of particles sorted in size classes and particle types according to their size in µm. You can filter by particle type and by approval status. Further tables are available to inform you about the test results in respect of length and width classes. See |
|
2 |
Scatter Plot Displays the particle values for length versus width. You can filter by particle type and by approval status. See Scatter Plot. |
|
3 |
Particle Button bar Filters artifacts or particles by type. You can view all particles clicking the All button. See Particle Button Bar and Filtering Particles. |
|
4 |
Image Gallery Displays all detected particles. Default: Minimized view. The Mid-Sized view shows the particle images with an enlarged zoom level. See TCA Image Gallery. |