Configuring a ZEN Data Storage Archive
The archive options enable you to specify how and where the Archive is stored, as well as the hierarchy of attributes that can be assigned to templates.
The Archive Setup button needs to be clicked once for the first client to set up the database. On the second client, this is not necessary anymore, even though the button is active.
- You are logged in as an administrator in ZEN core.
- You have configured a ZEN Data Storage Server and activated the Data Storage Client module in Home Screen > Maintenance > Toolkit Manager.
- Click Home Screen > Maintenance > Archive Options.
- The Archive Options dialog opens.
- As Archive Type, select ZEN Data Storage Archive from the drop-down list.
- Only if you have selected ZEN Storage Archive as archive type, you can setup the Archive.
- For Settings > URL specify the URL of the ZEN Data Storage server.
- The Archive Setup button is displayed.
- If you want to use an archive hierarchy, create your hierarchy elements by clicking +, see Creating a hierarchy for the archive and Using a hierarchy for the archive for general information.
- Click Archive Setup. This has to be done only on the first system which connects to the ZEN Data Storage.
- The storage is set up to work with the software and the standard users are created.
- Restart the software to update the user database.
- You have set up the ZEN Data Storage Archive as your archive and the database is set up properly. You can now configure the user management, see Basics of User and Group Management.