Acquiring Images - Best Practice
Technical Cleanliness Analysis is a standard based application, this means that the minimum technical requirements, hardware and software settings shall be fulfilled. In order to receive results which are mainly operator independent, the initial hardware and software settings shall be pre-defined by the supervisor in the following way.
For Technical Cleanliness Analysis with Axiocam 705 pol two Polarizer are mandatory. The Axiocam 705 is one Polarizer and the second polarizer must be placed in the microscope light path as well. Both directions of polarization must be 90° to each other.
To set up the system, the following is performed:
- The desired job template and the standards to be analyzed are selected, see Standard Template Editor.
- The Tiles (measurement area) workbench is selected, see Tiles (measurement area) Workbench.
- The live image is well focused and shows a mean image brightness.
- In the Extended Camera tool, the B/W color mode of the camera is selected. In dual camera mode, the color mode RGB is selected, see Extended Camera Tool.
- To make sure that the single tiles are oriented correct to each other, the camera-stage alignment is performed.
On the Tiles (measurement area) workbench, perform the following steps:
ABOUT ZEISS
Impressum
Carl-Zeiss-Strasse 22
73447 Oberkochen
Germany
Legal