Tiles Setup (manual) Tool
With this tool you can manually setup a tile region wether by entering the number of tiles (e.g. 3 x 3) or size of the tile region (e.g. 4000µm x 4000µm). The individual images (single tiles) are acquired and then stitched together to create the final tile image.
If the sample height varies by a large amount, we recommend to add a focus correction method. For this, add the Focus Correction tool and select a focus correction method. This is applied during the acquisition of the tile image.
|
Parameter |
Description |
|---|---|
|
Create tile region |
Creates a preview of the tile region which you have setup. The tile region is created at the current stage position. The yellow grid shows the tiles to be acquired, including an overlap. The frame shows the total (sample) area of interest. If you click on Start on top of the workbench, the final tile image will be acquired. |
|
Tiles |
If selected, you can enter the number of tiles to be acquired, e.g. 3x3 tiles. The size of the tile region is calculated automatically. |
|
Size |
If selected, you can enter the size of the tile region to be acquired in µm, e.g. 4000 x 40000 µm. The number of tiles is calculated automatically. |