Understanding the Illig Method
The Illig Workflow generates a joint result from previously performed job runs and consists of two phases:
- Preparation phase:
Generates results for the Illig workflow, see steps 1-4 below. - Illig calculation phase:
Summarizes several job results for calculation of the Illig value, see step 5 below.
- Select Job Mode > job template with VDA 19.1 or ISO 16232 Method Extended Analysis.
NOTICE! Don't modify the length size classes because the Illig calculation uses a predefined size classification. - Load the input form Technical Cleanliness (Illig Value).
NOTICE! The sample location and sedimentation time are automatically read-out and applied by executing the Illig job template Technical Cleanliness (VDA 19.2 Joint Result). - Run the pre-configured Component Cleanliness job template to analyze your particle traps.
- Repeat steps 1-3 until all particle traps are analyzed.
- Perform calculation of the Illig value by using the job template Technical Cleanliness (VDA 19.2 Joint Result).