VDA 19.2 Technical Cleanliness in Assembly - Environment, Logistics, Personnel and Assembly Equipment
The Illig method, as described in the VDA 19.2 standard, is used to test the cleanliness of a certain location with its environmental conditions, e.g., air or working benches. By means of particle traps the amount of sedimented particles per time unit (Illig Value) can be analyzed.
The detected particle number per size class is multiplied by a weighting factor, see the first table, summed up and normalized to calculate the Illig value, see table Illig weighting factors. The sum value is normalized to an area of 1000 cm2 and related to a measuring time of 1 h. The result is the Illig Value. The calculated Illig value creates a comparison basis for the collected particulate contamination at different locations over a certain time periode. With the Illig formular, larger particles are stronger weighted than smaller ones, because it is more likely that the larger ones have a higher damage potential.
|
Particle Size Class |
Size x in µm |
Weighting Factor |
|---|---|---|
|
B |
5 ≤ x < 15 |
0 |
|
C |
15 ≤ x < 25 |
0 |
|
D |
25 ≤ x < 50 |
0 |
|
E |
50 ≤ x < 100 |
1 |
|
F |
100 ≤ x < 150 |
4 |
|
G |
150 ≤ x < 200 |
9 |
|
H |
200 ≤ x < 400 |
16 |
|
I |
400 ≤ x < 600 |
64 |
|
J |
600 ≤ x < 1.000 |
144 |
|
K |
1.000 ≤ x |
400 |
Example calculation: Illig weighting factors
|
Size x in µm |
Result |
Weighting Factor |
Weighted Particle No. |
|---|---|---|---|
|
5 ≤ x < 15 |
-- |
0 |
0 |
|
15 ≤ x < 25 |
-- |
0 |
0 |
|
25 ≤ x < 50 |
1620 |
0 |
0 |
|
50 ≤ x < 100 |
374 |
1 |
374 |
|
100 ≤ x < 150 |
57 |
4 |
228 |
|
150 ≤ x < 200 |
43 |
9 |
387 |
|
200 ≤ x < 400 |
15 |
16 |
240 |
|
400 ≤ x < 600 |
7 |
64 |
448 |
|
600 ≤ x < 1.000 |
2 |
144 |
288 |
|
1.000 ≤ x |
3 |
400 |
1200 |
|
Result: |
3165 |
||
|
Normalized for 1000 cm2 and 1 h × 0.39 * |
1234 |
||
|
Illig Value [1/1000] cm2 h |
|||
Applying the Illig formular
1h / measuring time [h] × 1000 cm2/ measuring area [cm2] = 0.39
Time of sedimentation: 1 week = 168 h
Measuring area (лr2) 15.2 cm2
Measuring area: Filter membrane area used for analysis.
Sedimentation time [h]: Defined the time frame of the sample exposure to air.