To check the calibration-setup of your microscope system for a component cleanliness analysis, you can use a test slide (round glass slide) with a defined number of different objects. For this, ZEN core offers the following two test standards:
- Particle Test Standard (CC) (2010) Slide D45
- Particle Test Standard (CC with OC) (2010) Slide D45
The two standards are available in Manage Templates. You can acquire images of the test slide and use the standards together with the image analysis to determine if your microscope is configured as needed for the component analysis. To get a good and quick impression of the system status, we recommend to only image one group of objects on the slide.
The following example shows an acquisition of the test slide with two polarization channels and the segmentation result with the detected objects inside the drawn measurement frame:
|
Test slide with Pol 0 channel |
Test slide with Pol 90 channel |
|
Segmentation result with detected objects inside the measurement frame |
For accessing the information of your analysis, the table(s), report document(s) and particle documents are saved in Browse Results, and can be opened from there.
The following table shows an example result for Metallic Shiny Particles found on the test slide. The particle counting results are based on a measurement area with a diameter of 14.5 mm.