Before you run the Particle Test Standard to check the calibration, carry out the following preparations.
- Moisten the slide carefully with a soaked, soft and lint-free cloth (water/detergent solution). Clean the slide afterwards in fresh water. Gently dry the slide with a second dry, soft and lint-free cloth.
- We recommend to insert a plain white CN-Derivate filter membrane on the holder underneath the particle test standard. Place the slide with the particles shimmering yellowish on top of the plain filter membrane in the slide holder and turn the top lid carefully until it is fixed.
- Mount the filter holder in the stage insert and set a focus on the sample with the large cross in the center.
- You have prepared the calibration check using the test slide. If your Job Template is configured, continue with the check, see Routinely Measuring and Interpreting Results.