Configure the job template so that you can run the test standard on a regular basis.
- In the Light Path Editing tool, define an appropriate acquisition setting.
- Use the 10x objective magnification for the measurement.
- Adjust the exposure time best suited for acquisition of the POL-90 image and bright image POL-0, and use the determined exposure time for further measurement.
- Apply the shading correction on a particle free area, e.g. close to the cross in the center.
- Define a Tile Region with a diameter of 14.5 mm with the cross in the center.
- In the Frame Setup tool, set the parameter for the mode to Inside only.
- Use the SW-Autofocus for the measurement. A good starting parameter is to focus on each 25th tile.