This workflow description explains the workflow steps including the Switch workbench to select an active branch, as well as groups and loops. Here, you configure the workflow for the operator who acquires the images. The first steps in the workflow are the same, but from the second Switch workbench on, you configure the workflow. However, the last steps are again the same for the different workflows.

Technical Cleanliness Standard Selection

Provides a list of available standards for the TCA module that shall be used for the analysis.

  1. Load the available standards for the job template, see Technical Cleanliness Standard Selection workbench.
  2. Select the standards that are relevant for the operator's analysis, see Technical Cleanliness Standard Selection Tool.

NOTICE! For particle height measurement and correlative analysis only one standard can be selected.

Group

Group for Input Form loop, Tiles Measurement Area and Split Image workbenches.

The Define Outputs tool does not display any entries. Therefore, editing is not possible.

Loop

Loop for Input Forms.

Iteration of all working steps within the loop definition.

  1. Configure the number of loops the tasks in the next loop will perform.
  2. Sets the working steps for the operator, see Loop Task.

Input Form

In the Form Selection tool, input forms are provided.

  1. In the Create Form Image tool, configure the size of the input form image in the report.
  2. Select from the set of pre-defined input form templates an appropriate input form. For component cleanliness one input form per extraction method is provided and furthermore one general form with minimum content.
  3. Add the customer data to the corresponding file. The Operator selects the customer data.

The customers are managed with an Excel file. This file is stored per default here: C\ProgramData\Carl Zeiss\ZENcore\Customer.Contacts.xlsx

The form is displayed in the report, see Form Selection Tool and Create Form Image Tool.

Tiles (measurement area)

Configures the settings of the specimen acquisition.

  1. Configure the image acquisition and save it as individual setting in the Camera tool.
  2. Perform a shading correction.
  3. Adjust tile regions to the specimen area to be analyzed, and save it as individual setting in the Camera tool.
  4. Verify the focus points

See

The operator uses this settings for the specimen acquisition.

Split Image

Splits n tile region images into n individual overview images.

Default: Run silent

We recommend not to change the default parameters.

See Split by Dimension Tool.

Loop

Iteration of all working steps within the loop definition.

  1. Select the number or range of loop iterations. The default value is Batch processing, i.e. the number of iterations is determined by the setting in the first loop.

We recommend not to change the default values.

The operator will not see these settings, see Loop Settings Tool.

Image Processing

Changes the image format to 8 bit.

Default: Run silent

We recommend not to change the default parameters.

See Convert Pixel Format Tool.

Image Channel Mapping

Determines the channel allocation for image analysis.

Default: Run silent

No parameters are accessible.

See Image Channel Mapping Tool.

TCA Analysis Task Group (Component)

Provides the image analysis setting for component cleanliness.

The Technical Cleanliness Analysis setting is applied by default to determine the particle types, i.e. metallic-shiny, non-shiny, and fiber.

We recommend not to change the default parameters.

The operator will not see these settings.

See Load Setting Tool

Frame Setup

Sets up and adds a measurement frame to the image. Only the area inside the frame is used for the analysis. The area of the measurement frame is retrieved from the Tiles (measurement area) tool.

Default: Run silent

Activated: Uses Tile Region Geometry, i.e. the measurement frame is retrieved automatically from the configured tiles regions, see

Particle Segmentation (Components)

Extracts objects from the background by applying the pre-defined relative threshold range.

  1. Adjust the Relative Threshold if required. The Relative Threshold refers to the Relative Image Brightness (Luminosity).

Methods defined by the standards for component cleanliness:

  • Standard Analysis
    The threshold setting shall not be modified.
  • Extended Analysis - Free Analysis
    The threshold setting can be adjusted if required.

See Concept of Relative Image Brightness and Relative Threshold.

The operator uses these values for the image segmentation, see Particle Segmentation Tool (Components).

Region Filter

Applies a filter to Image Analysis results.

Default: Run silent. Feret Maximum is activated and set to a minimum of 5 µm for component cleanliness and to 1 µm for oil cleanliness.

Note: Only relevant for oil job templates: the default region filter setting in all oil job templates is by default 1 µm for the minimum particle size and required for SAE AS 4059. This setting increases the analysis time. For oil standards with a size classification starting from 5 µm the region filter setting might be adapted accordingly.

In the Image Gallery, all particles are displayed from 5 µm length onwards for component cleanliness testing and 1 µm length onwards for oil cleanliness testing.

Measurement Data

Shows measurement results for the detected particles after segmentation and image analysis.

Default: Run silent

The Measurement Data tool displays the particle results. The calculation is performed with default values in the background, see Measurement Data Tool.

The operator will not see these settings.

Standard Specific Settings

Configures specific parameter for the standard based calculation.

Default: Run silent

  1. Select the type of membrane filter in use.
  2. This determines the allowed range for the occupancy rate as defined by the standard.
  3. Adjust settings for the Normalization Parameter.
  4. Select whether the particle size class All shall display the results including fibers or not.
  5. Select how many of the largest particles per type you want to add to the report.
  6. The coordinates of all pre-selected particles will be added to the position list with EDF.
  7. Optionally, you can add the Equivalent Spherical Method (Components) tool to the workbench. The operator can adjust the Measurement Frame Area of the Frame Setup tool.
  1. These settings are conducted automatically in the operator workflow.

See

  • Occupancy Rate (Components) Tool
  • Normalization Parameter (Components) Tool
  • Type Classification for Class "All" Tool
  • Cleanliness Classification (Length) Tool
  • Particle Gallery (Report) Tool
  • Equivalent Particle Area (Components) Tool

Switch

Changes between the branches.

  1. Select the active branch according to the operator's needs, see Configuring the TCA Job Template by Switching the Active Branch.
    You have the following options:
    Size Distribution Live. See Size Distribution Live - Branch
    Size Distribution Live & EDF/Position List. This is a prerequisite for the selection of the height measurement branch. See Size Distribution Live EDF/Position List - Branch