These workflow steps are only visible for the active branch Size Distribution Live & EDF/Position List.
Size Distribution Live & EDF/Position List
Displays a group of tasks for workflow steps demanding hardware control.
The corresponding task group is loaded.
No further action required.
See Configuring the TCA Job Template by Switching the Active Branch.
Size Distribution (Live)
These workflow steps are only visible for the active branch Size Distribution Live.
No further action required.
See:
- Size Distribution View
- Standard Result Selection Tool
- Displayed Normalization Tool
- Cleanliness Classification (Length) Tool
- Specimen Overview Tool
- Statistical Analysis Tool
- Result Approval Tool
- Occupancy Rate (Components) Tool or Occupancy Rate (Oil, Lubricants) Tool
Switch
Changes between the branches.
- Size Distribution Live & EDF/Position List branch is selected.
- Select the branch according to the operator's needs.
You have the following options:
With Height Measurement
Without Height Measurement
For information on the branch switches, see Configuring the TCA Job Template by Switching the Active Branch.
Interactive Height Measurement
Only visible with active branch With Height Measurement.
Enables to perform interactive height measurements.
To perform the interactive particle height measurement conduct manual focusing as described:
- Manually define the z-position at the membrane filter beside the selected particle as lowest position.
- Manually define the z-position at the top of the particle as highest position.
- The absolute difference between both z-values is the determined particle height.
See Interactive Particle Height Measurement Tool.
Position List with EDF
If zero particles are found, for example, in a blank test, this step is skipped.
An EDF image is created.
Acquires automatically an EDF image of the largest particle per type for the particle gallery in the report. The corresponding particle coordinates are saved by default in the Position List tool. The displayed number of particles per type cannot be modified in the EDF step. Note: The total number of particles per type is defined in the Standard Specific Settings workbench in the Particle Gallery (Report) (Components) tool.
Define the initial microscope and camera setting. Adjust the step size for the automated EDF acquisition.
See:
- Light Path Editing Tool
- EDF Setup (motorized focus) Tool
- Extended Camera Tool
- Position List Tool
- Select Active Camera
Reset Device Parameters
Changes the microscope hardware and camera settings. Restores the initial microscope and camera settings as defined in the Tiles (measurement area) workbench.
No further action required.
Split EDF Image
Splits the image scene into n images.
Default: Run silent
We recommend not to change the default parameters.
See Split by Dimension Tool.
To follow the workflow steps description, see Last Workflow Steps Independent of Active Branch.