These last workflow steps are the same for all active branches.

Technical Cleanliness Results Output

Configures the layout of the gallery images in the report document and enables HYDAC Process Data Extraction.

With Shuttle & Find: enables the export of an EDS file with particle positions.

We recommend not to change the default parameter for the gallery image layout.

  1. Add the file path to where the HYDAC parameter XML file is stored.

See:

  • HYDAC Data Extraction Tool
  • With Shuttle & Find, see EDX File Export Tool

For Oil Cleanliness

Reports

This oil cleanliness specific workbench creates a report document for each specimen.

All selected standards are considered automatically.

See:

For Component Cleanliness

TCA Report

This component cleanliness specific workbench creates a TCA report for each specimen.

  1. In the Template Selection tool, the default report template is pre-selected.
  2. In the Display Settings tool, select a particle counting result, one or more chart types, and one or more particle types. Select System Properties to have the system properties displayed in the report.
  3. In the Gallery Setting tool, set the image count per row.
  4. The TCA report is configured for the operator. Note: The total number of particles per type is defined in the Standard Specific Settings workbench in the Particle Gallery (Report) (Components) tool.

See

  • Template Selection Tool
  • Chart Selection Tool
  • Gallery Setting Tool