This view options tab is only visible if you have opened a Lightfield 4D image document.
|
Parameter |
Description |
|
|---|---|---|
|
Raw Data |
Displays the raw data image of the selected single lens. |
|
|
Selected Lens |
Only visible if Raw Data is selected. |
|
|
Volume Definition |
Only visible during live acquisition. |
|
|
Volume Color Coding |
Only visible if Volume Definition is selected. |
|
|
– |
Top |
Displays the color for image areas that are above focus. |
|
– |
Center |
Displays the color for image areas that are in focus. |
|
– |
Bottom |
Displays the color for image areas that are below focus. |
|
Lightfield 4D |
Only visible for fully acquired Lightfield 4D images and not during a live acquisition. |
|
|
Iterations |
Sets the number of iterations of a (constrained iterative) deconvolution. A higher value results in sharper objects but will also amplify noise and measurement errors. The available value range is between 5 and 50 iterations. |
|
|
Background Correction |
Sets the rolling ball radius to remove homogeneous background for better reconstruction. Larger values remove bigger structures, while smaller values also remove smaller structures and have a stronger effect on the image. The available value range is between 75 and 5 pixels. |
|
|
Total Variation Denoising |
Reduces negative effects of noise on the deconvolution by smoothing homogenous areas while preserving edges (sharp contrasts). The available value range is between 0 and 100%. |
|
|
Parameter Configuration |
|
|
|
– |
Store |
Stores the current parameter values. |
|
– |
Recall |
Recalls parameter values from storage, if available. |
|
– |
Reset |
Resets all parameters to factory defaults. |
|
Process Image |
Selects the image dimensions, for example Current Timepoint, to create a processed image with the currently selected parameter values. |
|
|
Create Processed Image |
Creates an image processed with the current parameter values. The processing always takes the currently displayed image. If you have an image of a multi-position experiment and deactivate the Scene checkbox on the Dimensions tab, all positions are displayed in the Image View and processed. The image is saved at the location set in the Lightfield 4D Processing function, see Lightfield 4D Processing. To change the path in this tab, you have to change the storage location in the processing function and execute the function once on the Processing tab. |
|