To enable communication between SmartSEM and ZEN, the RRemote Client has to be installed on the SEM system as well. After the ZEN software is installed, install the RRemote Client.

  1. You have read the chapter Configuration of Security Software attentively.
  2. Your system is connected to the Internet.
  3. You have opened the ZEISS Microscopy Installer with admin rights and you are logged in.
  4. You have selected (default setting) the online installation, see ZEISS Microscopy Installer manual.
  5. You have closed SmartSEM and EM Server.
  1. Open the Install tab.
  2. On the left, activate ZEN with the respective version number.
  3. The available components and tools for installation are displayed on the right side.
  4. On the right side, activate SmartSEM Remote API.
  5. Activate I agree to the Terms and Conditions of the selected software.
  6. Click Install.
  7. The download starts. The progress is displayed in the installer.
  8. After the download, the installation starts. Several installation dialogs are displayed, but no action is required. Wait until the different windows have disappeared. After the installation procedure you can check the configuration with the tool RConfigure (open from Windows Start menu).
  1. You have successfully installed the Remote Client on your SEM system. You can test the program with the Desktop-App SmartSEM API Test.