This table displays the data for the different measurement positions. Each position is displayed as a separate row. If a value could not be calculated for a specific position, the corresponding entry in the table is empty. A click on an entry highlights and selects the entire row and displays the corresponding data in the chart. The units for individual parameters are displayed as tooltips. A right click in the table opens a context specific menu, see Parameter Table Context Menu.

Parameter

Description

Ø

Activated: Calculates an average of the individual values. The average is displayed on the bottom of the table.


Visibility

Toggles the visibility of the respective spot.

ID

Displays the ID for the individual spot.

Color

Sets the color for the respective measurement point by clicking on the field and selecting a color form the dialog.

Name

Displays and sets the name for the individual position.

Count Rate

Displays the count rate in kHz.

CPM

Displays the CPM in kHz. The color of the value indicates the data quality, see Data Quality Criteria (CPM).

Number of Molecules

Displays the number of molecules found at the measurement position. It is calculated by 1/correlation amplitude, where the correlation amplitude is defined as value of the correlation curve at the smallest lag time.

Concentration

Only visible after using a 3D fit model.
Displays the concentration.

Surface Concentration

Only visible after using a 2D fit model.
Displays the value for the surface concentration in nmol/cm2.

Triplet State Fraction

Displays the triplet state fraction. One physical property of the utilized fluorophore. Dependent on intensity and if applicable oxygen. A triplet state of e.g. 16% would mean that 16% of all fluorophores are in dark state.

Chi2

Displays the Chi2 value, see also Fit Calculation. Chi2 is the deviation between the measured FCS curve and the fit curve. A value of 0 would mean a 100% accordance.

Diffusion Model

Displays which fit model was used.

Diffusion Time

Displays the diffusion time.

Diffusion Coefficient

Displays the diffusion coefficient.

Date

Displays the date and time of the experiment.