The automatic settings you make through these buttons affect the settings displayed in the Channels tool, which are used for acquisition, see Channels Tool.

Parameter

Description

Automatic

  • Sets power of Colibri-LEDs depending on wavelength and power of transmitted and reflected light sources to 10% or 3V.
  • The system tries to set the optimal resolution for the camera by adjusting the binning if supported. The resolution is calculated from the total magnification at the camera adapter (objective + optovar + adapter).
  • Sets the Exposure Time:
    • Transmitted light channel: 10 ms
    • Reflected light channel: 10 ms
    • Fluorescence channel: 150 ms
  • Sets the Auto Exposure to Off
  • Sets Intensity to the following:
    • Transmitted light channel: 70%
    • Reflected light channel: 70%
    • Fluorescence channel: 30%

Speed

  • Sets power of Colibri-LEDs 100% and power of transmitted and reflected light sources to 10% or 3V.
  • The system tries to set the optimal resolution for the camera by adjusting the binning if supported. The resolution is calculated from the total magnification at the camera adapter (objective + optovar + adapter).
  • Sets the Exposure Time:
    • Transmitted light channel: 3 ms
    • Reflected light channel: 3 ms
    • Fluorescence channel: 50 ms
  • Sets the Auto Exposure to Off
  • Sets the Intensity to the following:
    • Transmitted light channel: 30%
    • Reflected light channel: 30%
    • Fluorescence channel: 15%

Signal

  • Sets power of Colibri-LEDs 75% and power of transmitted and reflected light sources to 10% or 3V.
  • The system tries to set the optimal resolution for the camera by adjusting the binning if supported. The resolution is calculated from the total magnification at the camera adapter (objective + optovar + adapter).
  • Sets the Exposure Time:
    • Transmitted light channel: 20 ms
    • Reflected light channel: 20 ms
    • Fluorescence channel: 300 ms
  • Sets the Auto Exposure to Off
  • Sets the Intensity to the following:
    • Transmitted light channel: 100%
    • Reflected light channel: 100%
    • Fluorescence channel: 60%

Default

  • Sets power of Colibri-LEDs 10% and power of transmitted and reflected light sources to 10% or 3V.
  • The system tries to set the optimal resolution for the camera by adjusting the binning if supported. The resolution is calculated from the total magnification at the camera adapter (objective + optovar + adapter).
  • Sets the Exposure Time to 20ms.
  • Sets the Auto Exposure to Default
  • Sets the Intensity to 100%.

Current

NOTICE! If you changed hardware settings manually and do not want to lose them, click Current.

  • Sets power of Colibri-LEDs depending on wavelength and power of transmitted and reflected light sources to 10% or 3V.
  • The system tries to set the optimal resolution for the camera by adjusting the binning if supported. The resolution is calculated from the total magnification at the camera adapter (objective + optovar + adapter).
  • Transfers the following values from the current camera in use to the Channels tool:
    • Auto Exposure
    • Exposure Time
    • Intensity